Müller Voltage Amplifier MVA 10
Low Noise Amplifier with 10 MHz Bandwidth
For Thermocouples and Hydrophones


Applications:
This Amplifier is a reasonable and the ideal product for our MCT Thermocouples. Together with the 1 MHz filter, it has an excellent signal to noise ratio for the small signals of the type E or type K thermocouples.
A second application is together with our Müller-Platte Needle Probe or M60-1. The signal of these hydrophones are quite small in applications like ultrasound.. However, together with the MVA 10 the range of application extends to the complete field of ultrasound. The amplifier will be used with full 10 MHz bandwidth to measure signals in a plane or focused field as well as for sinus or pulsed waves.
Other applications are general preamplifier for oscilloscopes and transient recorders, photomultiplyer, signal booster for optical receivers and current amplifiers.

image
image

image
1 MHz filter for connecting the output cable to the recorder

Technical Data
Input Impedance:1 MOhm, 15 pF
Input Voltage Noise:4.7 nV/√Hz at 2 MHz
Input Noise:100 µV (peak – peak)
Output Impedance:50 Ohm
Output Voltage:± 3,5 V
Max. Output Current:100 mA
Gain:x100 or x1000, switchable
Gain Accuracy:± 0.2 dB
Bandwidth DC:1 Hz to 10 MHz
Rise Time:35 ns
Supply Voltage:± 15 V, power supply is included in the kit
Supply Current:± 70 mA
Weight:200 g
Housing Material:Al/Mg4.5Mn, nickel plated
Operating Temperature:0 – 60°C
Connectors:BNC neg. for input and output
Absolute Max. Input Voltage :± 5 V
 
 
Ordering information
Article-No.: 200-120-1Amplifier MVA 10: 10 MHz Voltage Amplifier incl. Power Supply
Article-No.: 200-120-2Low pass filter 1 MHz
Dr. Müller Instruments
Hasengarten 35
61440 Oberursel
Germany

Tel.: +49 - 6172 - 380 37 27
Fax: +49 - 6172 - 177 077 4
Mobil: +49 - 171 - 431 9060